FAQ-4148

Do I have to wait for EZ2 Connect Fx to cool down before starting a new run?

No, one doesn’t have to wait for the cooldown of the device for a re-run with Sep&Prep. When starting a run to process non-sperm or any other samples with the EZ1&2 DNA Investigator Kit shortly after an Investigator Sep&Prep run, the heating block of the instrument is still at high temperature. This may cause increased evaporation of elution buffer, which is transferred to the heating position after a protocol is started, resulting in slightly lower eluate volumes. If this is observed, wait until the heating block has cooled to a lower temperature. Note that the elution buffer is heated to 65°C during a run. Thus, cooling to lower temperatures is not required.