FAQ-2160

Which elution volumes can be selected using the QIAsymphony Virus Blood 200 protocol?

Final elution volumes of 60 µl, 85 µl, 110 µl, or 165 µl can be selected using the QIAsymphony Virus Blood 200 protocol. Please note that the initial elution volumes are 95 µl, 120 µl, 145 µl, or 200 µl. When calculating the amount of internal control(s) as well as the titer of the processed sample, it is necessary to take into consideration the initial volume of elution buffer that is used for each sample.

 