QIAxcel DNA Kits

QIAxcel 기기를 사용한 DNA 절편 자동 분석에 사용됩니다

S_1084_5_GEN_V2

✓ 24/7 automatic processing of online orders

✓ Knowledgeable and professional Product & Technical Support

✓ Fast and reliable (re)-ordering

QIAxcel DNA High Resolution Kit (1200)

Cat. No. / ID:  929002

QIAxcel DNA High Resolution Gel Cartridge, 완충액, 미네랄 오일, QX Intensity Calibration Marker, 12-튜브 스트립
Copy order details
₩1,715,000.00
Log in To see your account pricing.
Cartridge type
High Resolution
Screening
Fast Analysis
High Sensitivity
QIAxcel DNA Kits은/는 분자생물학 분야에 사용하기 위한 것입니다. 이 제품들은 질병의 진단, 예방, 또는 치료 목적으로 사용할 수 없습니다.

✓ 24/7 automatic processing of online orders

✓ Knowledgeable and professional Product & Technical Support

✓ Fast and reliable (re)-ordering

Features

  • 즉시 사용 가능한 겔 카트리지로 안전하고 편리함
  • 간소화된 워크플로우로 분석 시간 최소화
  • 5pg/µl의 낮은 핵산 농도에서도 확실한 결과 제공
  • 3–5bp까지의 분해능으로 정확한 분석
  • 디지털 데이터 수집을 사용한 표준화된 처리

Product Details

QIAxcel 기기를 QIAxcel DNA Kits와 함께 사용하면 실행당 최대 96개의 샘플에서 DNA 절편의 크기 분리와 정량화가 완전 자동으로 수행됩니다. QIAxcel 기술은 겔 카트리지를 사용한 모세관 전기영동을 기반으로 비교할 수 없는 분해능, 속도, 처리량을 제공합니다. QIAxcel 겔 카트리지는 재사용이 가능하여 샘플 12개를 250회까지 실행할 수 있습니다. 대부분의 응용 분야에 적합한 사전 설치된 방법이 제공됩니다. QIAGEN은 사용자를 위한 맞춤형 방법을 생성할 수도 있습니다.

QIAxcel DNA High Sensitivity Kit는 QIAxcel Connect에서만 작동합니다. 기타 모든 QIAxcel DNA Kits는 QIAxcel Connect 또는 QIAxcel Advanced 기기와 함께 작동합니다. QIAxcel DNA Accessories는 모든 QIAxcel DNA Kits와 호환됩니다.

Performance

간편한 고분해능 분석

유기체 유전형 분석, PCR 기반 cDNA 라이브러리 선별 및 기타 수많은 생물의학 연구 및 분자생물학 응용 분야에서는 소수의 염기쌍으로 다양한 크기의 PCR 절편을 식별해야 합니다. QIAxcel High Resolution Kit를 QIAxcel 기기와 함께 사용하면 15bp-5kb 크기의 DNA 절편을 분석할 수 있습니다. 500bp보다 작은 절편은 3–5bp의 분해능으로 분리할 수 있어 데이터 해석 측면에서 기존의 고분해능 아가로스 겔 전기영동법보다 향상된 신뢰도를 제공합니다( 암 관련 돌연변이 유전자의 고감도 검출 참조). QIAxcel 기기는 희석되지 않은 증폭 반응에서 0.1ng/µl DNA의 검출 민감도로 더 적은 양의 샘플 투입 물질을 사용하여 더 강력한 결과를 제공합니다.

QIAxcel DNA Screening Kit는 제한 엔도뉴클레아제 분해, 정성적 싱글플렉스 또는 멀티플렉스 PCR 및 기타 DNA 응용 분야에 사용할 절편의 일상적인 평가에 매우 적합합니다( 멀티플렉스 PCR 샘플의 고분해능 분석 참조). DNA 선별에 겔 카트리지를 사용하면 크기가 15bp-5kb인 절편의 분석 시간이 최소화되어(샘플 96개가 약 1시간 안에 처리됨) 실험실 워크플로우가 간소화됩니다. 1kb 미만의 절편에서 20–50bp의 분해능을 얻을 수 있습니다. 샘플 소비량이 0.1µl 미만이므로 귀중한 샘플을 거의 모두 이후의 다운스트림에 사용할 수 있습니다.

QIAxcel DNA Fast Analysis Kit는 정성적 싱글플렉스 또는 멀티플레스 PCR 응용 분야에서 DNA 절편의 일상적인 평가에 매우 적합합니다. DNA 선별에 겔 카트리지를 사용하면 크기가 15bp—3kb인 절편의 분석 시간이 최소화되어(샘플 12개가 단 3분 만에 처리되고 96개는 약 25분 안에 처리됨) 실험실 워크플로우가 간소화됩니다( PCR 절편 분석 데이터를 보여주는 스크린샷 참조). QIAxcel DNA Fast Analysis 겔 카트리지를 사용하면 50bp까지의 분해능을 얻을 수 있으며, 샘플 사용량은 0.1μl 미만입니다.

QIAxcel DNA High Sensitivity Kit를 사용하면 5pg/μl의 낮은 농도에서도 DNA를 분석할 수 있습니다. 크기가 15bp–3kb인 절편의 분석 시간은 샘플 12개의 경우 약 9분입니다. QIAxcel DNA High Sensitivity 겔 카트리지를 사용하면 20–50bp의 분해능을 얻을 수 있습니다.

See figures

Principle

QIAxcel 기기는 모세관 겔 전기영동법을 사용하여 핵산을 크기별로 빠르게 분리합니다( 샘플 분리 과정 참조). 기존의 아가로스 겔 전기영동법과 달리 즉시 사용 가능한 겔 카트리지의 모세관에서 분리가 수행됩니다. 샘플이 자동으로 개별 모세관에 로드되고 전압이 적용됩니다. 음전하를 띤 핵산 분자가 모세혈관을 통해 양전하를 띤 말단으로 이동합니다. 저분자량 분자는 아가로스 겔 전기영동법과 마찬가지로 고분자량 분자보다 빠르게 이동합니다. 분자가 모세관을 통해 이동하면서 형광 신호를 감지 및 측정하는 검출기를 통과합니다. 광전 증폭관은 형광 신호를 전자 데이터로 변환한 다음 QIAxcel ScreenGel 소프트웨어를 사용하여 추가 처리를 위해 컴퓨터로 전송합니다. 처리가 끝나면 데이터가 전기영동도 또는 겔 이미지로 표시됩니다.

See figures

Procedure

QIAxcel DNA Kits는 DNA 절편의 분리 및 분석을 단순화하고 사용자의 안전을 개선합니다. 몇 가지 간단한 단계만 수행하면 됩니다. 즉, 겔 카트리지를 로드하고, 완충액 트레이를 채워 로드하고, 96-well 플레이트나 PCR 튜브 또는 스트립에 샘플을 로드하고, 사용할 프로세스 프로필을 선택하기만 하면 됩니다. 지루한 겔 준비와 광범위한 사용자 교육이 필요하지 않아 워크플로우를 간소화하고 시스템을 일상적인 실험실 루틴에 쉽게 통합할 수 있습니다. 실행을 시작한 후 몇 분 내에 컴퓨터 화면에 첫 번째 결과가 실시간으로 나타납니다( 실행 시간 단축 참조).

See figures

Applications

해당 겔 카트리지와 함께 사전 프로그래밍된 방법을 사용하면 싱글플렉스 또는 멀티플렉스 PCR 절편, 제한 엔도뉴클레아제로 분해한 DNA, 합성 올리고뉴클레오티드를 포함하여 다양한 핵산을 분리하고 분석할 수 있습니다.

  QIAxcel DNA High Resolution Kit QIAxcel DNA High Sensitivity Kit§ QIAxcel DNA Screening Kit QIAxcel DNA Fast Analysis Kit
크기 범위 15bp–20kb, 최대 50kb 표시 15bp–3kb 15bp–5kb 15bp–3kb
분해능* 3–5bp 20–50bp 20–50bp 50bp
검출 한계 0.1ng/μl 5pg/μl 0.1ng/μl 0.1ng/μl
카트리지당 샘플 샘플 1200개 샘플 1200개 샘플 2400개 샘플 3000개
실행 시간 9–25분 약 9분 약 7분 약 3분
응용 분야 영역
  • 고분해능 유전형
  • VNTR, STR 등
  • gDNA 무결성 검사
  • NGS 라이브러리 검사
  • 액체 생검 추출물의 품질 관리
  • NGS 라이브러리 검사
  • 신속 PCR 선별(싱글플렉스 및 멀티플렉스)
  • 제한 분해, 플라스미드
  • gDNA 무결성 검사
  • NGS 라이브러리 검사
  • 신속한 단일 앰플리콘 분석(예: HLA 형별 검사)

Supporting data and figures

Resources

Brochures & Guides (5)

For analysis and typing of PCR products from different HLA loci using the QIAxcel Advanced and the Helmberg-SCORE software

A protocol for evaluation of samples for next-generation sequencing on the QIAGEN GeneReader platform using the QIAxcel Advanced Instrument and ScreenGel software version 1.5 of higher
Kit Handbooks (2)
For automated analysis of DNA fragments using the QIAxcel and QIAxcel Advanced instruments
For automated electrophoretic analysis of DNA using the QIAxcel Connect instrument

June 2022
Application Notes (20)
The QIAxcel System was successfully used to detect DNA derived from genetically modified organisms (GMOs) at a level suitable for GMO testing according to EU standards.
Operating Software (2)
This release includes the first version of QIAxcel ScreenGel NGS Profiles. These profiles are intended to be used to run and analyze reactions of GeneRead QIAact AIT and BRCA 1/2 panels using the QIAxcel instrument and QIAxcel ScreenGel 1.5 Software. These profiles are composed of an Process Profile, Analysis Profile, Distribution Profile, and Report/Export Profile.
Quick-Start Protocols (1)
For electrophoretic analysis of DNA samples using the QIAxcel Connect system

June 2022
Instrument User Manuals (2)
For use with QIAxcel ScreenGel software v2.0
For use with QIAxcel Advanced instruments and QIAxcel ScreenGel Software version 1.6
Certificates of Analysis (1)

FAQ

Why does an intensity calibration of the gel cartridge for the QIAxcel System and QIAxcel Advanced need to be performed?

The intensity calibration procedure on the QIAxcel System and QIAxcel Advanced is performed to normalize the signal intensities across all 12 channels of the gel cartridge. This corrects for natural intensity reading variations between each capillary in the cartridge. A correction factor is determined and applied for every subsequent run performed with the cartridge.

FAQ ID -1824
Can a QIAxcel DNA Fast Analysis Cartridge be run on a QIAxcel System?

The QIAxcel DNA Fast Analysis Cartridge can only be run on the QIAxcel Fast Analysis System with the BioCalculator Fast Analysis software and the grey software key.

All other QIAxcel Cartridges can only be run on the QIAxcel System with the BioCalculator software and the blue software key.

FAQ ID -1828
Can customers use their own DNA size marker with the QIAxcel System and if yes, what concentration should be used?

The concentration of the DNA marker, pUC18/HaeIII, provided in the QIAxcel DNA Kits is 5ug/ml in 1XTE buffer solution. Customers can run their own DNA size marker sample with different concentration levels using the provided Methods (i.e. M500, H500, L500 and so on).

The recommended method for all basic applications with the QIAxcel System is the M500 method. The BioCalculator software also allows customers to choose their own Sample Injection Time in the Sequence table. The M500 method provides 5KV for 20 seconds for the electro-kinetic injection of lower concentrated DNA samples. If the Injection Time is doubled or tripled (40 or 60 sec), you can expect double or triple intensities (also peak heights).

FAQ ID -1841
Where are the QIAxcel data stored?

For the BioCalculator software, the default data folder is:

C:\Program Files\QIAGEN\QIAxcel BioCalculator\Data.

For the ScreenGel software, the default data folders are:

C:\Documents and Settings\All Users\Application Data\QIAGEN\QIAxcel\ScreenGel\Data\Experiment (for a computer with Windows XP operation system) or C:\ProgramData\QIAGEN\QIAxcel\ScreenGel\Data\Experiment\ (for a computer with Windows 7 operating system).

Alternatively, you can more conveniently access the ScreenGel experiment data folder using the menu options File\Open data directory\Experiment within the ScreenGel software.

FAQ -9008
If a different DNA size and/or alignment marker is to be used with the QIAxcel System or QIAxcel Advanced, does the Gel Cartridge need to be recalibrated?

The Gel Cartridge does not need to be recalibrated when using a new alignment marker and/or DNA size marker with the QIAxcel System or QIAxcel Advanced. Calibration is required only once, i.e. when the first cartridge is run or when the cartridge is used on a different QIAxcel system than the one it has been calibrated on, or when using a different PC than the one used to calibrate initially.

FAQ ID -1858
How can drying out of the gel cartridge tips used on the QIAxcel System and QIAxcel Advanced be prevented?

If the gel cartridge tips for the QIAxcel System and QIAxcel Advanced are in contact with air, the tips will dry out. There are 3 ways of preventing the tips from drying.

  1. Store the cartridge in the instrument with the instrument in the “Park” position. The washing solution (8 ml) covered with mineral oil must be in the “Park” position of the Solution Tray. 
  2. Store the cartridge in the Cartridge Stand with enough mineral oil in the well of the Cartridge Stand to cover the capillary tips. The Cartridge Stand should only be used for short term storage, i.e. when changing cartridges during one working day. 
  3. Place the cartridge back into its packaging with the capillary tips gently inserted into the soft gel in the box.
FAQ ID -1823
Why does the ScreenGel software version 1.2 crash when I try to use an old reference size marker table?

The reference size marker table created by ScreenGel version 1.1 or earlier is incompatible with ScreenGel version 1.2. Re-analyze the size marker run in ScreenGel version 1.2 and create a new reference size marker table for future use.

FAQ -9009
Why is the cartridge not recognized?

When using the BioCalculator software, ensure the color of the software key matches the type of cartridge used. For example, the QIAxcel Fast Analysis cartridge requires the red software key, while the QIAxcel High Resolution cartridge requires the blue software key.

In addition, check that the communication between the computer and the instrument is established. Further, check that the smart key is inserted securely.

If the issue persists, contact QIAGEN Technical Services and provide the cartridge ID and lot number, which is printed on the packaging of QIAxcel cartridges.

FAQ -9004
How do I perform the filter test?

When using the BioCalculator software, select Settings under File in the Instrument control panel window. In the new windows, click Check Filter in the new window and follow the on-screen prompts.

When using the ScreenGel software, log in to the ScreenGel software with an account with basic or advanced privilege.  Select System Check under the Service environment. Select Filter Check and click Start.

Please send the filter test result to QIAGEN Technical Services for further assistance.

FAQ -9006
Why is the signal of the QX alignment marker weak?

Incorrect storage can lead to QX alignment marker degradation. The QX alignment markers can be used for 50 runs or 3 days, whichever comes first. Opened QX alignment marker should be store at –20°C between uses.

FAQ -9003
What causes the calibration to fail?

Clogged capillaries and/or re-using the calibration marker are the most common causes of failed calibration. Perform “Removing gel from blocked capillaries”, as outlined in Appendix D in the QIAxcel DNA handbook. Use a new set of calibration marker for each new calibration procedure.

FAQ -9001
I have upgraded the ScreenGel software to version 1.2. How can I stop the software crash when I try to select a cartridge?

Please delete the 6 obsolete method files as outlines in the Important Note QIAxcel ScreenGel Software

FAQ -9010
How do I perform the detector test?

When using the BioCalculator software, select Detector test under File in the Instrument control panel window. Perform the Detector test with and without a cartridge latched inside the QIAxcel. Feedback is not shown on the screen when you select the Detect test. Instead, the file named detector.asc is saved at C:\Program Files\QIAGEN\QIAxcel BioCalculator. Change the file name before performing another detector test.

Note: Older versions of BioCalculator software save the detector.asc at C:\Program Files\QIAxcel BioCalculator.

When using the ScreenGel software, log in to the ScreenGel with an account with basic or advanced privilege.  Select System Check under the Service environment. Select Detector Test and click Start. The file is saved at C:\Documents and Settings\All Users\Application Data\QIAGEN\QIAxcel\ScreenGel\Data\Troubleshooting\Troubleshooting_xxxxxxxx (current date in the format of YYYYMMDD) for a computer with Windows XP operating system or C:\ProgramData\QIAGEN\QIAxcel\ScreenGel\Data\Troubleshooting for a computer with Windows 7 operating system. Alternatively, you can perform a screen capture for the Detector test result.

Please send the Detector test results to QIAGEN Technical Services for further assistance.