PyroMark Q24

Managing instrument methods in PyroMark Q24 Software

PyroMark instruments dispense the correct volume of enzyme mix, substrate mix, and nucleotides using variable pulse times and dispensing pressures. The settings for these pulse times and dispensing pressures are specific for each lot of PyroMark Q24 Cartridges and correspond to the method number printed on the PyroMark Q24 Cartridge. The instrument method should be selected in the PyroMark Q24 Software according to the reagents and reagent cartridge that will be used for the run. The settings and instrument method files provided on this page are only for use with PyroMark Q24 Software and PyroMark Gold Q24 Reagents.

 Integrated Automated Solutions, Pyromark Q24
Important: The settings and instrument method files provided on this page should only be used for Pyrosequencing analyses set up with PyroMark Q24 Software and PyroMark Gold Q24 Reagents.
  • Access the available downloadable instrument method files below or under the "Resources" tab on the PyroMark Q24 product page.
  • Click the instrument method file that corresponds to the method number printed on the label of the cartridge to be used.
  • Save the instrument method file to the computer running PyroMark Q24 Software.
  • In the PyroMark Q24 Software, open the "Instrument Method" dialog box from the "Tools" menu.
  • Click the "Import" button and locate the downloaded instrument method file.
  • The downloaded instrument method file will now appear in the "Instrument Methods" dialog box under "Methods provided by user" and can be selected in the dropdown list when creating a new Run Setup.
  • Open the "Instrument Method" dialog box from the "Tools" menu.
  • Click an existing instrument method, e.g., "PyroMark Q24 Method 001 Rev. A created 2008-04-23" which is already implemented in the PyroMark Q24 Software 2.0.
  • In the table below, find the dispensing pressure and pulse time settings that correspond to the method number printed on the label of the cartridge to be used.
  • Enter the dispensing pressures and pulse times in the corresponding fields of the "Instrument Method" dialog box.
  • Optional: enter information in the notes field, such as file name and date.
  • Click "Save As" and save the method with a new name (e.g., PyroMark Q24 Method 0003).
  • When creating a new Run Setup, select this new method (e.g., PyroMark Q24 Method 0003) from the instrument method dropdown list.

Instrument methods for various lots of PyroMark Q24 Cartridges (cat. no. 979202) for the PyroMark Q24 instrument

  Dispensing pressures (mbar) Pulse times (ms)
Method Nucleotides Reagents Enzyme mix Substrate mix Nucleotides
0001 625 400 49 45 4.4
0002 625 400 55 50 4.4
0003 625 400 63 57 4.4
0004 625 400 63 57 4.2
0005 625 400 62 57 4.3
0006 625 400 62 57 4.5
0007 625 400 62 57 4.0
0008 625 400 57 49 4.5
0009 625 400 57 49 4.0
0010 625 400 60 54 4.0
0011 625 400 62 57 4.5
0012 625 400 62 57 4.5
0013 625 400 62 55 4.5
0014 625 400 62 56 4.5
0015 625 400 60 54 4.5