For easy setup of the protocol to evaluate pipetting accuracy of the EZ1 Advanced XL

    The EZ1 Advanced XL Test Card provides protocols designed to evaluate the pipetting accuracy of the EZ1 Advanced XL. The EZ1 Advanced XL pipets either 100 µl or 500 µl of water from one set of tubes to another. Pipetting accuracy should be within defined specifications of ±8 µl for 100 µl, corresponding to pipetted volumes in the range of 92–108 µl and ±40 µl for 500 µl, corresponding to pipetted volumes in the range of 460–540 µl. An additional protocol on the EZ1 Advanced XL Test Card describes how to test the tip adapters for liquid leaks.

    Product Product no. Cat. no. List price:
     
     
    show details
        varies
    Can't order online?
    To place an order via phone, email or for requesting a quote, please provide the product’s name, number and catalog number.
    Product Cat. no. List price:
    EZ1 Advanced XL Test Card
    Preprogrammed card for the EZ1 Advanced XL test protocol
    9018706
    $844.00

    EZ1 Advanced XL Cards are used in combination with EZ1 Kits on the EZ1 Advanced XL. Please refer to the specific EZ1 Kit handbook and the EZ1 Advanced XL User Manual for information about product use.