✓ 24/7 automatic processing of online orders
✓ Knowledgeable and professional Product & Technical Support
✓ Fast and reliable (re)-ordering
Cat. No. / ID: 9016187
✓ 24/7 automatic processing of online orders
✓ Knowledgeable and professional Product & Technical Support
✓ Fast and reliable (re)-ordering
EZ1 Instrument Test Cards provide protocols designed to evaluate the pipetting accuracy of EZ1 instruments. The EZ1 Advanced XL, EZ1 Advanced or BioRobot EZ1 pipets either 100 µl or 500 µl of water from one set of tubes to another. Pipetting accuracy should be within defined specifications of ±8 µl for 100 µl, corresponding to pipetted volumes in the range of 92–108 µl and ±40 µl for 500 µl, corresponding to pipetted volumes in the range of 460–540 µl. An additional protocol on EZ1 Instrument Test Cards describes how to test the tip adapters for liquid leaks.