For reliable screening of changes in methylation status by HRM analysis
  • Accurate detection of changes in CpG methylation status
  • Highly specific PCR amplification without the need for optimization
  • Distinct melting curves due to saturating EvaGreen fluorescent dye
  • Convenient master mix format and optimized protocol
  • Fast and easy development of new HRM methylation assays
The EpiTect HRM PCR Kit uses high-resolution melting (HRM) technology for fast screening and accurate detection of changes in the CpG methylation status of bisulfite converted DNA. The optimized kit is available in a convenient master mix format and contains the novel double-stranded DNA-binding fluorescent dye EvaGreen, as well as an optimized HRM buffer, HotStarTaq Plus DNA Polymerase, and dNTPs. The kit is compatible with real-time instruments suitable for HRM analysis, particularly the Rotor-Gene Q, and also the Rotor-Gene 6000 and the LightCycler 480.
Product Product no. Cat. no. List price:
 
 
show details
    varies
Can't order online?
To place an order via phone, email or for requesting a quote, please provide the product’s name, number and catalog number.
Product Cat. no. List price:
EpiTect HRM PCR Kit (100)
For 100 reactions: 2x EpiTect HRM Master Mix containing HotStarTaq Plus DNA Polymerase, EpiTect HRM PCR Buffer (with EvaGreen dye), dNTP Mix, and RNase-Free Water
59445
$204.00
The EpiTect HRM PCR Kit is intended for molecular biology applications. This product is not intended for the diagnosis, prevention, or treatment of a disease.
  • Main Image Navi
Highly sensitive results.|Methylation analysis on various real-time PCR instruments.|Principle of methylation analysis of the APC promoter by HRM.|Principle of HRM technology.|
Mixtures of methylated and unmethylated DNA of varying ratios were used as template. A CpG island from the promoter region of the APC gene (adenomatosis polyposis coli) was amplified, and the degree of methylation was determined by HRM methylation analysis on the Rotor-Gene Q using the EpiTect HRM PCR Kit. [A] A standard normalized melt curve and [B] a difference plot normalized to the 50% methylated sample are shown. Even low percentages of methylated DNA are detected.|Mixtures of methylated (M) and unmethylated (U) DNA in various ratios were used as template. A CpG island from the promoter region of the PTPN6 gene (protein tyrosine phosphatase, nonreceptor type 6) was amplified, and the degree of methylation was determined by HRM methylation analysis. [A] Difference plot when using the EpiTect HRM PCR Kit and the Rotor-Gene Q. [B] Difference plot and melting peaks of 100% unmethylated and methylated DNA, respectively (inset) when using the EpiTect HRM PCR Kit and the LightCycler 480. [C] Difference plot and melting peaks of 100% unmethylated and methylated DNA, respectively (inset) when using Supplier R's HRM kit (with 3.5 mM MgCl2) on the LightCycler 480. The kit was tested with 1.5 mM and 3.5 mM MgCl2 (within the lower and upper limits of the recommended Mg2+ concentration). Amplification with 1.5 mM MgCl2 failed (data not shown); amplification with 3.5 mM MgCl2 resulted in the generation of nonspecific products when using methylated DNA as template (red double peaks), limiting the HRM analysis.|[A] Genomic DNA — all cytosines that are not part of CpG sites are symbolized by X. CpG sites containing potentially methylated cytosines are indicated as CG. [B] Bisulfite conversion of genomic DNA — all unmethylated cytosine residues are converted to uracil. Cytosines within CpG sites are indicated by Y and are either converted to U (if unmethylated) or remain unmodified (if methylated).|HRM analysis is based on the dissociation behavior of dsDNA due to increasing temperature. Melting of dsDNA depends on GC-content. AT-rich regions melt faster.|
Performance
Novel fluorescent dye EvaGreen

EvaGreen is a dsDNA-binding dye. The spectral properties of EvaGreen are similar to those of SYBR® Green I. The absorbance maximum is 500 nm, and the emission maximum is 530 nm. In contrast to SYBR® Green I, EvaGreen can be used at higher, saturating concentrations without PCR inhibition and shows equal binding affinity for GC-rich and AT-rich regions with no apparent sequence preference. This makes EvaGreen an ideal dye for HRM analyses of all types of PCR product.

Highly sensitive detection of changes in CpG methylation status

For CpG methylation analysis by HRM, the DNA has to undergo complete bisulfite conversion before the melting behavior of DNA can be measured by HRM analysis. Unknown samples can be compared with a standard control, based on their melting characteristics. The sensitivity ensured by the EpiTect HRM PCR Kit means that even low amounts of methylated DNA can be detected (see figure Highly sensitive results).

Consistent performance on various real-time instruments

The EpiTect HRM PCR Kit delivers highly accurate methylation results with real-time instruments suitable for HRM (see figure "Methylation analysis on various real-time PCR instruments"). The kit includes an optimized protocol that has been tested on the Rotor-Gene Q, Rotor-Gene 6000, and LightCycler 480. The combination of advanced HRM technology on the Rotor-Gene Q with the benefits of the EpiTect HRM PCR Kit delivers exceptional HRM analysis.

Principle

HRM is an innovative technique that characterizes double-stranded PCR products based on their dissociation (melting) behavior as they transition from double-stranded DNA (dsDNA) to single-stranded DNA (ssDNA) with increasing temperature. Amplification of bisulfite-converted DNA is carried out using primers that are conversion-specific but not methylation-specific. The primers should therefore comprise several converted cytosines and should flank CpG sites (see figure "Principle of methylation analysis of the APC promoter by HRM"). This is to ensure amplification of bisulfite-converted DNA only and enable distinction between methylated and unmethylated CpG sites during HRM analysis.

The fluorescence of EvaGreen is measured continuously and is plotted against increasing temperature. EvaGreen is only measured as long as it is bound to dsDNA. This results in high fluorescence at the start of HRM analysis. As the increasing temperature gives rise to DNA melting, EvaGreen is released and the fluorescence reduces to a background level. GC-rich stretches of DNA are more stable and therefore melt more slowly compared with AT-rich regions and remain double stranded at higher temperatures for longer (see figure "Principle of HRM technology").

HRM analysis is an easy and cost-effective alternative to probe-based methylation assays. PCR products can be discriminated according to sequence, length, GC content, or strand complementarity — down to single base pair changes.

Convenient master mix format

The EpiTect HRM PCR Kit is provided in a convenient master mix format for increased ease-of-use. The optimized 2x EpiTect HRM PCR Master Mix ensures highly specific amplification. In addition, it enables flexible, rapid, and sensitive analysis of methylation status of CpG dinucleotides in bisulfite converted DNA via HRM. The master mix consists of optimized concentrations of HotStarTaq Plus DNA Polymerase, EpiTect HRM PCR Buffer, dNTPs, and EvaGreen dye.

HotStarTaq Plus DNA Polymerase

HotStarTaq Plus DNA Polymerase, a modified form of QIAGEN Taq DNA Polymerase, is supplied in an inactive state that has no polymerase activity at ambient temperature. This prevents extension of nonspecifically annealed primers and primer dimers formed at low temperatures during PCR setup and the initial PCR cycle.

Optimized EpiTect HRM PCR Buffer

EpiTect HRM PCR Buffer maintains specific amplification in every cycle of PCR. The balanced combination of KCl and (NH4)2SO4 in the buffer promotes specific primer–template annealing. Simultaneously, nonspecific annealing is reduced, maximizing yields of specific PCR product, which allows optimal HRM analysis. Specific amplification with high product yield is maintained without the need for time-consuming optimization of the Mg2+ concentration.

Procedure

The target sequence is PCR amplified to a high copy number in the presence of the dsDNA binding fluorescent dye, EvaGreen. The dye does not interact with single-stranded DNA (ssDNA), but actively binds to dsDNA and fluoresces brightly when bound. Following the amplification, the temperature is increased, and the fluorescence of EvaGreen is measured continuously and plotted against the temperature.

Applications

The EpiTect HRM PCR Kit provides a highly reliable and cost-effective method for methylation analysis using HRM technology and can be successfully used for methylation studies in epigenetic research. The optimized protocol means that information on the methylation status of CpG sites can be derived quickly and efficiently. This simplifies methylation assay development.